Pulsatile mock loop systems are largely used to investigate the cardiovascular system in vitro. They consist of a pump, which replicates the heart, coupled with a lumped-parameter hydraulic afterload, which simulates vasculature. An accurate dimensioning of components is required for a reliable mimicking of the physiopathological behavior of the system. However, it is not possible to create a component for the afterload inertance, and inertance contributions are present in the entire circuit. Hence, in the literature, in-ertance is neglected or qualitatively evaluated. In this paper, we propose two quantitative methods (Maximum-likelihood estimation (MLE) and Bayesian estimation) for estimating afterload inertance based on observed pressure and flow waveforms. These methods are also applied to a real mock loop system. Results show that the system has an inertance comparable with the literature reference value of the entire systemic circulation, and that the expected variations over inlet average flow and pulse frequency are in general confirmed. Comparing the methods, the Bayesian approach results in higher and more stable estimations than the classical MLE.

(2013). Inertance estimation in a lumped-parameter hydraulic simulator of human circulation [journal article - articolo]. In JOURNAL OF BIOMECHANICAL ENGINEERING. Retrieved from http://hdl.handle.net/10446/170483

Inertance estimation in a lumped-parameter hydraulic simulator of human circulation

Lanzarone, Ettore;
2013-01-01

Abstract

Pulsatile mock loop systems are largely used to investigate the cardiovascular system in vitro. They consist of a pump, which replicates the heart, coupled with a lumped-parameter hydraulic afterload, which simulates vasculature. An accurate dimensioning of components is required for a reliable mimicking of the physiopathological behavior of the system. However, it is not possible to create a component for the afterload inertance, and inertance contributions are present in the entire circuit. Hence, in the literature, in-ertance is neglected or qualitatively evaluated. In this paper, we propose two quantitative methods (Maximum-likelihood estimation (MLE) and Bayesian estimation) for estimating afterload inertance based on observed pressure and flow waveforms. These methods are also applied to a real mock loop system. Results show that the system has an inertance comparable with the literature reference value of the entire systemic circulation, and that the expected variations over inlet average flow and pulse frequency are in general confirmed. Comparing the methods, the Bayesian approach results in higher and more stable estimations than the classical MLE.
articolo
2013
Lanzarone, Ettore; Ruggeri, Fabrizio
(2013). Inertance estimation in a lumped-parameter hydraulic simulator of human circulation [journal article - articolo]. In JOURNAL OF BIOMECHANICAL ENGINEERING. Retrieved from http://hdl.handle.net/10446/170483
File allegato/i alla scheda:
File Dimensione del file Formato  
Paper.pdf

Solo gestori di archivio

Versione: publisher's version - versione editoriale
Licenza: Licenza default Aisberg
Dimensione del file 2.08 MB
Formato Adobe PDF
2.08 MB Adobe PDF   Visualizza/Apri
Pubblicazioni consigliate

Aisberg ©2008 Servizi bibliotecari, Università degli studi di Bergamo | Terms of use/Condizioni di utilizzo

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10446/170483
Citazioni
  • Scopus 11
  • ???jsp.display-item.citation.isi??? 11
social impact